|本期目录/Table of Contents|

[1]张焰,周宇△,侯健,等.基于双频并行模型的小肠组织压榨状态研究*[J].生物医学工程研究,2019,04:450-454.
 ZHANG Yan,ZHOU Yu,HOU Jian,et al.Study on the compression state of small intestine tissue based on dual-frequency parallel model[J].Journal of Biomedical Engineering Research,2019,04:450-454.
点击复制

基于双频并行模型的小肠组织压榨状态研究*(PDF)

《生物医学工程研究》[ISSN:1006-6977/CN:61-1281/TN]

期数:
2019年04期
页码:
450-454
栏目:
出版日期:
2019-12-25

文章信息/Info

Title:
Study on the compression state of small intestine tissue based on dual-frequency parallel model
文章编号:
1672-6278 (2019)04-0450-05
作者:
张焰周宇△侯健李媛陈琳宋成利
上海理工大学医疗器械与食品学院,上海 200082
Author(s):
ZHANG YanZHOU YuHOU JianLI YuanCHEN LinSONG Chengli
University of Shanghai for Science and Technology, Shanghai 200082,China
关键词:
吻合器自动压榨双频并行模型ColeY模型压榨状态
Keywords:
Stapler Automatic pressing Dual-frequency parallel model ColeY model Squeezed state
分类号:
R318
DOI:
10.19529/j.cnki.1672-6278.2019.04.14
文献标识码:
A
摘要:
基于自制的自动压榨装置对胃肠组织的压榨状态进行研究,采用双频并行阻抗分析的方法来研究压榨过程中组织阻抗的变化情况,探索可以表征组织受压状态的压榨参数,以期用于智能吻合器的控制。模拟吻合器的工作方法,控制恒压强分别为8、16、32 g/mm2对小肠组织进行压榨,测量组织阻抗进行分析。结果显示,细胞低频电导GLF和高频电纳BHF随着压强的迅速增加而减小;恒压时,GLF随时间增加而减少;压强较小时,BHF随时间波动减小;压强较大时,BHF的波动减弱,减小趋势增加。双频并行模型参数GLF和BHF的变化趋势与基于ColeY模型的多频阻抗分析中G0、CCPEF和△G的变化趋势具有相似性,因此,也可以作为反映细胞受压状态的工程化参数。
Abstract:
To study the pressing state of gastrointestinal tissue based on self-made automatic pressing device. We studied on the change of tissue impedance during pressing process by dual-frequency parallel impedance analysis,explored the pressing parameters that characterized the stressed state of the tissue for later control of the smart stapler,and simulated the working method of the stapler, to control the constant pressure of 8, 16 and 32 g/mm2 for the small intestine tissue, measure the tissue impedance for analysis. The results showed that the cell low-frequency conductance GLF and high-frequency susceptance BHF decreased with the rapid increase of pressure; At constant pressure, GLFdecreased with time; BHFdecreased fluctuation with time, When the pressure is large, the fluctuation of BHF was weakened and the decreasing trend was increased. The variation trend of dual-frequency parallel model parameters GLF and BHF was similar to the trend of G0,CCPEF and△G in multi-frequency impedance analysis based on ColeY model. So it may be used as an engineering parameter reflecting the state of the cell pressure.

参考文献/References

备注/Memo

备注/Memo:
(收稿日期:2019-06-20)上海市科研计划项目(18441900200)。△通信作者Email:zhouyu_working@163.com
更新日期/Last Update: 2020-01-03