|本期目录/Table of Contents|

[1]李浩,于华杰,李胜,等.拉曼光谱技术在鉴别瑞氏-姬姆萨染色的肿瘤细胞与正常细胞中的应用*[J].生物医学工程研究,2020,03:309-313.
 LI Hao,YU Huajie,LI Sheng,et al.Application of Raman spectroscopy in the identification of Wright-Giemsa stained tumor cells and normal cells[J].Journal of Biomedical Engineering Research,2020,03:309-313.
点击复制

拉曼光谱技术在鉴别瑞氏-姬姆萨染色的肿瘤细胞与正常细胞中的应用*(PDF)

《生物医学工程研究》[ISSN:1006-6977/CN:61-1281/TN]

期数:
2020年03期
页码:
309-313
栏目:
出版日期:
2020-09-25

文章信息/Info

Title:
Application of Raman spectroscopy in the identification of Wright-Giemsa stained tumor cells and normal cells
文章编号:
1672-6278(2020)03-309-05
作者:
李浩1于华杰23李胜124韩露2沈阳洋2△
1.山东省药物研究院,济南 250062;2.山东省肿瘤防治研究院,济南 250117;3.济南大学医学与生命科学学院,济南 250220;4.山东省医学科学院,济南 250062
Author(s):
LI Hao1YU Huajie23LI Sheng124HAN Lu2SHEN Yangyang2
1.Shandong Pharmaceutical Research Institute, Jinan 250062, China;2.Shangdong Cancer Hospital and Institute,Jinan 250117;3.School of Medicine and Life Sciences,University of Jinan,Jinan 250220;4.Shandong Academy of Medical Sciences, Jinan 250062
关键词:
拉曼光谱技术拉曼光谱特征峰 瑞氏-姬姆萨染色肿瘤细胞正常细胞
Keywords:
Raman spectroscopy Raman spectral characteristic peakWrightGiemsa stainingTumor cellsNormal cell
分类号:
R318
DOI:
10.19529/j.cnki.1672-6278.2020.03.16
文献标识码:
A
摘要:
为探究利用显微共聚焦拉曼光谱仪辨别经瑞氏-姬姆萨染色后的肿瘤细胞与正常细胞的可行性。应用拉曼光谱技术在532 nm波长激光下分别检测经瑞氏-姬姆萨染色技术染色的肿瘤细胞与正常细胞,对比6种肿瘤细胞、正常细胞和瑞氏-姬姆萨染料的拉曼光谱特征峰值,正常细胞和肿瘤细胞在208~215 cm-1、278~285 cm-1、336~341 cm-1峰位均有拉曼特征峰。发现6种肿瘤细胞的拉曼光谱峰值和瑞氏-姬姆萨染料的拉曼信号峰值均相同。经分析认为,检测到的均为各种染料的拉曼信号而非6种肿瘤细胞的拉曼信号。因此,无法用显微共聚焦拉曼光谱仪区分经瑞氏-姬姆萨染色后的肿瘤细胞与正常细胞。
Abstract:
To explore the feasibility of using micro confocal Raman spectrometer to discriminate tumor cells and normal cells stained by Wright-Giemsa dye,we used Raman spectroscopy to detect the tumor cells and normal cells with Wright-Giemsa stain under 532 nm wavelength laser, and compared the Raman spectral characteristic peaks of 6 kinds of tumor cells, normal cells and Wright-Giemsa dye.We found that the Raman peaks of 6 kinds tumor cells were the same as those of Wright-Giemsa dye. The results showed that the detected Raman signals were the Ramam signals of Wright-Giemsa dye, not those of 6 kinds tumor cells. Therefore, it is impossible to discriminate tumor cells and normal cells by using micro confocal Raman spectrometer.