[1]李红梅△,张素文,朱肖杰,等.顶空气相色谱法测定交联透明质酸钠凝胶中异丙醇的残留量[J].生物医学工程研究,2015,03:190-191.
LI Hongmei,ZHANG Suwen,ZHU Xiaojie,et al.Determination of the Residual Isopropanol in Cross-linked Sodium Hyaluronate by Gas Chromatography[J].Journal of Biomedical Engineering Research,2015,03:190-191.
点击复制
顶空气相色谱法测定交联透明质酸钠凝胶中异丙醇的残留量(PDF)
《生物医学工程研究》[ISSN:1006-6977/CN:61-1281/TN]
- 期数:
-
2015年03期
- 页码:
-
190-191
- 栏目:
-
- 出版日期:
-
2015-09-25
文章信息/Info
- Title:
-
Determination of the Residual Isopropanol in Cross-linked Sodium Hyaluronate by Gas Chromatography
- 文章编号:
-
1672-6278 (2015)03-0190-02
- 作者:
-
李红梅△; 张素文; 朱肖杰; 郈秀菊; 孟建文; 王晓晨; 朱小敏
-
山东省药学科学院, 山东省医用高分子材料重点实验室,济南 250101
- Author(s):
-
LI Hongmei; ZHANG Suwen; ZHU Xiaojie; GOU Xiuju; MENG Jianwen; WANG Xiaochen; ZHU Xiaomin
-
Shandong Academy of Pharmaceutical Science,Shandong Provincial Key Laboratory of Biomedical Polymers,Jinan 250101,China
-
- 关键词:
-
交联透明质酸钠凝胶; 顶空气相色谱法; 异丙醇残留量; 交联; 检测
- Keywords:
-
Cross-linked Sodium hyaluronate; Headspace gas chromatography; Residual isopropanol; Crosslinkingl; Determination
- 分类号:
-
R318
- DOI:
-
-
- 文献标识码:
-
A
- 摘要:
-
建立了气相色谱法测定交联透明质酸钠凝胶中异丙醇残留量的方法。选用DB-624毛细管柱(30 mm×0.535 mm,3.0 μm),氢火焰离子化检测器(FID),载气为氮气,流速为2.0 ml/min,柱温采用程序升温,初始温度为50℃,保持2 min,以10℃/min升温至160℃,后运行温度:50℃,保持1 min。异丙醇在浓度0.01083~0.8664 mg/ml的范围内,峰面积和浓度呈现良好的线性关系,相关系数0.9990;平均回收率为99.1%,相对标准偏差(RSD)为1.2%。该方法重现性好、灵敏度高、简单、准确。
- Abstract:
-
A method for determination of resiual isopropanol in cross-linked sodium hyaluronate by GC chromatography was established.The method was programmed with Agilent DB-624 capillary column(30 mm×0.535mm,3.0 μm)and FID detector.The carrier was nitrogen and the flow rate was 2.0 ml/min,the oven introduced sequential increasing of temperature programming.The initial column temperature was 50℃ and maintained it for 2 min,then raised it to 160℃ at a rate of 10℃/min,after running temperature to 50℃,maintained it for 1 min.The calibration showed a good calibration linearity with the range of 0.01083~0.8664 mg/ml for isopropanol and the correlation coefficient was 0.9990.The average recoveries were in the range of 99.1%.The relative standard deviations(RSD) was 1.2% .The method has good repeatability and high sensitivity and the method is simple and accurate.
备注/Memo
- 备注/Memo:
-
(收稿日期:2015-07-02)通信作者Email:yijianmei2003@163.com